Chapter 2, Operation; Test Sequence; Alarm Test; Gas Test - MSA QuickCheck Altair 4 Manual De Instrucciones

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Chapter 2,
Operation
1. Verify that tubing is connected and power is supplied.
2. Verify that automatic regulator is ON.
The calibration cylinder concentrations must match
the instrument calibration setting; otherwise, instrument
readings may be incorrect.
3. Insert instrument into the test station.
NOTE: The following test sequence occurs:

Test Sequence

Alarm Test

Audible alarm and visual alarm functions are checked
ALARM TEST LED lights
LED flashes amber during test
Green LED indicates passing
Red LED indicates failure
Solid amber LED indicates that Altair 4 alarms
are turned OFF.

Gas Test

Sensor response to gases is verified
GAS TEST LED lights
LED flashes amber during test
Green LED indicates passing
Red LED indicates failure.

Overall Pass/Fail Test Indication

PASS OR FAIL LED lights
Green LED indicates passing
Red LED indicates failure
Checkmark symbol appears on the instrument display
when it is successfully bump checked
" WARNING
2-1