Sartorius PT-15 Instrucciones De Manejo página 7

Instrumento portátil para medir ph | mv | temperatura
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

Key Functions
1
4
6
7
Figure 3
1 and 2 [
] and [
temperature values. They have no effect on the instrument when operating in ATC mode.
Hold the UP [
] and the DOWN [
increase. When in the Setup screen, the UP [
to select either buffer set 4.00, 7.00, 10.00 or 4.01, 6.86, 9.18.
3 [SETUP] key: The [SETUP] key selects buffer.
4 [MODE] key: The [MODE] key selects the parameters to be displayed.
Pressing the [MODE] key changes the display sequentially to display pH and mV.
The calibration values will not be affected by changing the display modes.
5 [ENTER] key: Press [ENTER]; the large LCD will display the efficiency of the sensor.
Press and hold [ENTER] for 3 seconds and the meter will clear the calibration and reset all
factory settings.
6 [CAL] key: Press to calibrate.
7 [ON/OFF] key: The [ON/OFF] key turns the meter on or off.
Setup
Mode
Cal
] keys. The [
] and [
] keys for five seconds. The scrolling speed will
Enter
] keys are used to manually enter the
] and DOWN [
] keys will allow the user
2
3
5
7