3. Installation
Terminal assignment (rear panel)
CRISON INSTRUMENTS, S.A.
E-08328 ALELLA - Barcelona
Made in Spain
4-20 mA
Pt 100 / Pt 1000
VCA
DANGER
Relays:
- Limit the current up to 3 A by using one fuse.
- Do not connect cables with voltage higher than 24 V.
12
si629 P_HACH_LANGE_Ed.0510.indd ANG:12
si629 P_HACH_LANGE_Ed.0510.indd ANG:12
Transmitter si629 P
S/N:
IP 54
pH / mV
ELECTRODE
A
L1
L2 / C
Important notes
:
• The si629 P does not require ground connec-
tion (terminal 1).
• In environments with interference problems it
is recommended to connect the instrument to
ground in terminal 1.
It is also recommended to use shielding cable,
both for the electrode signal (triaxial cable) and
the temperature probe.
The external shield of the triaxial cable and the
probe cable must be connected to terminal 16.
• If the temperature probe has a 2-wire cable
instead of a 4-wire, fit a jumper between ter-
minals 12 and 13
and another one
between 14 and
15, as indicated in
the scheme.