Setting the limits
Temperature
When changing the temperature unit from °C to °F, the set limit
values are converted
Humidity
When changing the humidity unit (%, td, wb), the set limit values
are not converted. The limit values must be reset via the Setup
menu.
Pressure
It is not possible to set the limit values.
Performing settings
1. Press and hold
> Display flashes.
2. Select the required function with [Set].
> The adjustable parameter flashes.
3 Set values with
settings, see following table).
-
Instrument changes to the next display function
To scroll rapidly forward and back in the relevant menu,
press and hold
Press
[Esc]
Hold
[Set]
down for 2 s to save previously set values and
leave the menu.
If the entire Setup menu is run through, the settings made
are automatically stored.
If no key is pressed within 30 s, the screen reverts to the
current reading display automatically. No settings are
saved.
-