Example: when calibration data are exported at 15:12:25 (3:12:25 pm) the name of the file will be:
C151225.txt
Note
●
Pressing EXIT during the exporting aborts the process
4.13 Temperature compensation
We recommend the use of either a built-in or a separate temperature probe. If a temperature probe is
used, ATC and the sample temperature are displayed. If no temperature sensor is used, MTC is dis
played and the sample temperature should be entered manually.
The meter accepts both NTC 30 kΩ and PT1000 temperature sensors, with selection taking place
either automatically or manually (see "Menus and settings: Temperature settings").
In the pH mode, the meter calculates the temperature-adjusted electrode slope using this temperature
and shows the temperature-compensated pH/ion value in the measurement display.
20
Operating the meter