Technical Data
Pipettes:
1 to 100 mL pipettes, glass or plastic
Working and recharging temperature:
50 to 95°F (10 to 35°C)
Batteries:
Two replaceable 1.2V nickel metal hydride
(NiMH); approximately 8 hours operation
Charger:
65 mA/h for 14 hours
Materials
Housing:
ABS
Nose cone:
Polypropylene
Pipette holder:
Silicone
Filter:
PTFE 25mm dia 0.45µm filter;
Polypropylene housing
The Rota-Filler 3000 ® must not be used with liquids whose vapors
are incompatible with silicone, ABS+Polyoxymethylene, PTFE,
polypropylene, and NBR.
Warning!
This instrument must not be used or recharged in an explosive
atmosphere. Highly flammable liquids (e.g. ether, acetone, and
other liquids with a flash point below 0°C) must not be
pipetted.
This mark is the confirmation that the unit conforms
to the EU guidelines and has been tested according
to the Low Voltage Directive 73/23/EEC and
Electromagnetic Compatibility Directive 89/336/EEC.
Disposal Responsibilities:
The equipment you purchased may contain hazardous substances that could impact
the environment. Per regulations on electronic devices in the European Community, you
must use the appropriate disposal systems to avoid exposure of these substances to
the environment. The disposal systems will reuse or recycle hazardous materials from
your equipment responsibly.
The crossed-out wheeled bin symbol invites you to use those systems.
If you need more information on the collection, reuse and recycling systems, please
contact your local or regional waste administration. You can also contact us for more
information on the environmental performance of our products.
6
Getting Started
Remove the contents from the package, examining them carefully for
breakage, defects, or shortages. Save the packaging.
Charging the Batteries
For transport safety reasons, the Rota-Filler 3000 ® is shipped with
minimal battery charge. The batteries must be charged for 14 hours
before initial use.
Warning!
Use original battery charger only!
The use of a different charger could damage the batteries and
would void the warranty. Confirm that the voltage of the adapter
matches the voltage of the electrical outlet.
Before starting to charge, confirm that the voltage or the battery
adapter is identical to the main power supply at the outlet.
1. The temperature range for recharging is 50 to 95°F (10 to 35°C).
2. Plug the charger jack into the base of the Rota-Filler 3000 ® .
3. Insert the plug of the adapter into a suitable wall outlet. The red
indicator on the charger will light up.
4. Place the Rota-Filler 3000 ® on the benchtop stand.
5. Allow to charge 14 hours from empty to fully charged.
6. Once the Rota-Filler 3000 ® is fully charged, the adapter
automatically switches to a trickle charge operation and the red
indicator will turn off. If the charge is interrupted and then
re-started, the adapter will start the full 14 hour charging cycle
again. This will not damage the batteries.
Notes
— When fully charged, the Rota-Filler 3000 ® can be operated for
approximately 8 hours.
— If using the Rota-Filler 3000 ® daily, do not charge the battery until the
"low-battery indicator" LED has lit. A residual charge of approximately
10%
is available when the LED has lit.
—The Rota-Filler 3000 ® can be operated during recharging.
— If the LED on the adapter does not light up when the Rota-Filler 3000 ® is
connected to the adapter, the Rota-Filler 3000 ® and adapter should both
be tested and if necessary, repaired or replaced.
7