3) Pull the foam cover over the prosthesis.
4) Set the connecting element onto the footshell or prosthetic foot. Depend­
ing on the version, the connecting element engages in the edge or rests
on the foot adapter.
5) Install the prosthetic foot on the prosthesis.
6) Mark the outer contour of the connecting element on the distal face of
the foam cover.
7) Disassemble the prosthetic foot and remove the connecting element.
8) Clean the connecting element using a degreasing cleaner.
9) Glue the connecting element onto the distal face of the foam cover
according to the marked outer contour.
10) Let the glue dry (approx. 10 minutes).
11) Install the prosthetic foot and adapt the exterior cosmetic shape. Take
into account compression caused by cosmetic stockings or SuperSkin.
6 Cleaning
1) Clean the product with a damp, soft cloth.
2) Dry the product with a soft cloth.
3) Allow to air dry in order to remove residual moisture.
7 Maintenance
► A visual inspection and functional test of the prosthetic components
should be performed after the first 30 days of use.
► Inspect the entire prosthesis for wear during normal consultations.
► Conduct annual safety inspections.
8 Disposal
This product may not be disposed of with regular domestic waste in all juris­
dictions. Disposal that is not in accordance with the regulations of the coun­
try where the product is used may have a detrimental impact on health and
the environment. Please observe the information provided by the responsible
authorities in the country of use regarding return, collection and disposal
procedures.