Conductivity Temperature Display and Calibration

Conductivity Temperature Display
Star Plus meters allow the temperature to be viewed on individual measurement
lines in addition to the temperature display on the top, left of the screen .
1 . In the measurement mode, press
The arrow icon will point to the selected line .
2 . Press
/
be changed to display conductivity (µS/cm or mS/cm), total dissolved solids
(mg/L), salinity (ppt), resistivity (MΩ-cm), temperature (no icon) or a blank line .
Conductivity Temperature Calibration
The temperature calibration mode of the Star Plus meter allows the temperature
on each measurement line to be manually adjusted .
1 . In the measurement mode, press
and press
2 . Press
to begin the calibration .
3 . When the reading stabilizes, the arrow icon and the first digit will flash .
Enter the temperature by pressing
move to the next digit .
4 . Press
to save and end the calibration .
Thermo Scientific Orion Star
to change the value on the selected line . The middle line can
/
until the temperature is shown for the selected line .
and Star Plus Meter User Guide
to choose the middle display line .
to choose the middle measurement line
/
to adjust each digit and
Conductivity Technique
to
EN-43