Disposal
1. Open the module cover.
2. Squeeze the two external
locking
lugs
in
withdraw the sealing insert
downwards.
3. Release the hoses from
the sealing insert.
4. Replace the sealing insert.
5. Fit the hoses to the new
sealing insert (note the
arrows
on
the
sealing
insert and the hose).
6. Squeeze
the
external
locking lugs in and insert
the sealing insert into the
module cover until the
locking lugs engage.
7. Check that the sealing
insert is fitted correctly in
the module cover.
6.4
Changing the fresh air filter
NOTICE
i
i
With regular use and depending on the conditions of
use, the fresh air filter should be changed typically
every 2 months.
1. Unscrew the old fresh air filter.
2. Screw in the new fresh air filter.
24
and
6.5
Calibrate touchscreen
1. While starting the system, press and hold the function key
until the calibration display appears.
2. Press the five consecutively displayed position markers as
they appear.
6.6
Cleaning
CAUTION
!
Abrasive cleaning implements (brushes etc.), cleaning
agents and cleaning solvents can destroy the fresh-air
filter.
The device does not need any special care.
If very dirty, the equipment can be carefully wiped down
with a damp cloth.
Carefully dry the device using a cloth.
7
Disposal
Dispose of the product in accordance with the applicable rules
00633286.eps
and regulations.
Disposing of electric and electronic equipment:
In accordance with EU Directive 2002/96/EC this
product must not be disposed of as household waste.
This is indicated by the adjacent icon.
You can return this product to Dräger free of charge.