EN
Technical Specifications
DE
1. Dimensions: 2.4x1.5x1.3in
FR
(62.2x37x33.1mm)
ES
Weight: 1.5oz (42.5g)
IT
2. Peak wavelength range of the light
emitted from the sensor: red light
660 nm ± 3; infrared light 905 nm ± 5.
3. Maximum optical power output of
the sensor: 1.2 mW for infrared light
(905 nm).
4. Normal working conditions:
Working temperature
Relative humidity
Atmospheric pressure
Rated voltage
41°F to 104°F (5°C to 40°C)
15% to 80%, non-condensing
70 kPa to 106 kPa
DC 3.0 V
17
5. Default values for alerts:
Parameter
Value
Upper limit: 100
Oxygen saturation
Lower limit: 94
Upper limit: 130
Pulse rate
Lower limit: 50
When the alert is turned on
and the actual measured
Alert condition
value goes beyond the
preset alert parameter range,
the Oximeter will play an
alarm sound.
18
EN
DE
FR
ES
IT