NOTICE
Use under unallowable environmental conditions
Damage to product due to unallowable environmental conditions
► Do not expose the product to unallowable environmental conditions
(see the section "Environmental Conditions").
► If the product has been exposed to unallowable environmental condi
tions, check it for damage.
► If damage is apparent or in case of doubt, do not continue using the
product.
► Take suitable measures if required (e. g. cleaning, repair, replacement,
inspection by the manufacturer or a specialist workshop etc.).
► Inform the patient.
Signs of changes in or loss of functionality during use
Reduced spring effect or decreased forefoot resistance are indications of
loss of functionality.
4 Scope of Delivery
The following single components and accessories are included in the scope
of delivery in the specified quantities and are available for reordering accord
ing to their respective symbols as single components ( ), single compon
ents with minimum order quantities ( ) or single-component packs ( ):
Fig.
Item
–
–
–
–
–
–
–
–
–
–
–
–
5 Preparation for Use
CAUTION
Incorrect alignment or assembly
Injuries due to fall resulting from damaged prosthesis components
► Observe the alignment and assembly instructions.
14 | Ottobock
1C10 Terion
Qua
Designation
ntity
1 ( ) Instructions for Use
1 ( ) Prosthetic Foot
1 ( ) Footshell
1 ( ) Connection cap
1 ( ) Toe insert
1 ( ) Protective Sock
Reference num
ber
647G972
–
2C12=*
2C13=*
2Z661=*
SL=Spectra-Sock