4.9 Sealing Access to the Balance Settings
The Menu Lock switch prevents changes to the Lock Menu. The switch can be secured using paper seals, wire
seals or plastic ties.
Un-Locked
4.10 Approved "M" Models
Balances marked with an "M" have been calibrated and sealed at the factory for use as an EEC approved balance.
The Cal, Setup, Mode and Unit menus will be locked.
4.11 Local Approval
The InCal models can be approved by local weights and measures authorities. The calibration must be verified and
the Menu settings must be set appropriately before being locked using the Lock Menu. The Menu Lock switch on
the bottom of the balance can then be sealed in the locked position. The supplied capacity label is to be placed
below the display.
5.0 MAINTENANCE
5.1 Troubleshooting
Symptom
Cannot turn on
Poor accuracy
Cannot calibrate
Cannot access mode
Cannot access unit
Err 5
Err 7.0
Err 8.1
Err 8.2
Err 8.3
Err 8.4
Err 9.5
Err 9.8
Error 53
REF Err
LOWREF
5.2 Service Information
If the troubleshooting section does not resolve or describe your problem, contact Fisher Scientific.
2014_391_10889_2_Precision-Analytical Series.indb 11
Possible Cause
No power to balance
Improper calibration
Unstable environment
Unstable environment
Incorrect calibration masses
Mode not enabled
Units not enabled
Average Piece Weight too small
Time out
Pan has load during power on
Pan was removed prior to power on
Weight on pan exceeds capacity
Pan was removed during weighing
Factory calibration data corrupted
Factory calibration data corrupted
EEPROM Checksum error
Reference Weight is too small
Reference Weight is too low for accurate
parts counting or percent weighing.
www.eu.fishersci.com/fisherbrand
Locked with Plastic Tie
Remedy
Verify connections and voltage.
Perform calibration
Move balance to suitable location
Move the balance to suitable location
Use correct calibration masses
Enter menu and enable mode
Enter menu and enable units
Add additional samples
Remove weight from pan and re-zero.
Install pan and re-zero.
Remove weight from the pan
Re-install pan
Contact the authorized dealer
Perform calibration
Contact the authorized dealer
Add additional samples
Add additional samples or continue to weigh
with less accurate results.