CLEANING/STORAGE
TP472 T/PUMP

5.0 Cleaning

Clean outer surfaces of the T/Pump with a damp cloth
and soapy water.
To clean the fluid system, prepare a germicidal solution
according to the manufacturer's instructions. Use
GAYMAR catalog REF MTA33. Drain the pump. Fill the
reservoir to the operating level indicated on the side of
the pump. Set the temperature indicator to its lowest
setting (fully counterclockwise), start the T/Pump, and
circulate the solution for one hour. Drain the solution
and refill the pump with distilled water.
Change the distilled water monthly or more often
depending upon use.
6.0 Storage (Short term)
Disconnect pad. Connect ends of the connector
hose together. Open hose clamps. Leave water in the
reservoir. To prevent hose kinks, coil the hose rather
than folding it. Fasten the hose and cord with strap.
6.1 Storage (Long term)
Disconnect pad. Connect ends of the connector hose
together. Open hose clamps. Add
/
ounce GAYMAR
1
4
MTA33 germicidal concentrate to water already in
reservoir. Run for two (2) minutes. Drain pump. To
prevent hose kinks, coil the hose rather than folding it.
Fasten the hose and cord with strap and store pump.
6.2 Draining
Disconnect the T/Pump from power source. Disconnect
the pad or hoses from one another, keeping hoses at or
above the level of the T/Pump. Remove the fill cap and
invert the T/Pump over a sink. When all fluid has drained
from the hoses and reservoir, replace the fill cap and
connect the hose
ends together.