Access level
Typical rights
0
View parameters, change views
1
+ Start / Stop measurements
2
+ Calibration
3
+ Modify parameters
4
+ Modify table "User Access level" + Enable/Disable "Access right"
At startup all menus are locked and a valid ID and password combination is required to get access
beyond the standard measurement view. Refer to
Note: If the instrument security is enabled and the login credentials are not known, contact Hach Service support
with the recovery code to get the login credentials. The recovery code shows on the login window. The supplied
login credentials expire in one day. Make sure to change the login credential with known values.

7.1.3 User management

Select Access table from the Security menu to show the list of registered users (a maximum of
99 users allowed). They are listed by name, ID, password and access level.
Pressing on an empty line or the Add button displays a window to add a new user. User name, ID,
password (minimum 4 characters) and access level (1 to 4) are required.
Pressing on a registered user displays a window for editing or deleting that user.

7.2 View menu

7.2.1 Numeric view

This is the default view. Display shows the numeric measurement value identified for each gas
measurement channel used, a graphic showing measurement value evolution during the set time
frame, and sample temperature. The display is refreshed after each measurement cycle ( based on
the oxygen level for LDO sensor). This display can be configured to suit individual conditions and
convenience.
The K-M1100 low range sensors measure dissolved oxygen up to a maximum value of 5000 ppb.
Below 2000 ppb the measurement cycle interval is . Between 2000 and 3000 ppb the measurement
cycle interval is 30 seconds. Above 3000 ppb the measurement cycle interval is 60 seconds. The K-
M1100 high range sensors measure dissolved oxygen up to a maximum value of 40 ppm, the
measurement cycle interval is . Should the measured concentration go above the maximum value for
the sensor then the measurement cycle is increased to 60 seconds and an Out of range message
will be displayed. An arrow symbol to the right indicates if the value is increasing, decreasing, or
remaining constant. Once the measured value falls below the maximum value, the measurement
cycle returns to the pre-defined interval.

7.2.2 Numeric view configuration

1. Select Configure from the View menu followed by Conf. numeric view to customize the display:
Option
Display temperature
Display channel 1, 2, 3
Display mini graph
Display time base
Upper bound
Lower bound
Time base
Grid button
18 English
Start up
Description
Select Channel x to display the sample temperature for that channel.
Select yes or no
Check the box to display the graph.
Check the box to display the time base.
Adjust graph upper limit.
Adjust graph lower limit.
Adjust graph time span.
Set up the graph to display the x or y axes, grid, or alarm thresholds.
on page 17.