4 General safety instructions
CAUTION
Reaching into the area of the joint mechanism
Pinching of limbs (e.g. fingers) and the skin due to uncontrolled
joint movement
► Do not reach into the joint mechanism during daily use.
► Close attention is required during assembly and adjustment
tasks.
CAUTION
Mechanical damage to the product
Risk of injury due to change in or loss of functionality
► Use caution when working with the product.
► If the product is damaged, check it for proper function and readi
ness for use.
► In case of changes in or loss of functionality, do not continue
using the product (see "Signs of changes in or loss of functional
ity during use" in this section).
► Take any necessary measures (e.g. repair, replacement, inspec
tion by the manufacturer's customer service, etc.).
Signs of changes in or loss of functionality during use
Changes in functionality can manifest themselves, for example,
through a changed gait (swing phase, stance phase), incomplete
extension, stiffness and the development of noise.
5 Scope of delivery
The scope of delivery is shown in Figure
Only product components with reference numbers in the illustration
can be reordered separately.
1
on page 2.
6 Technical data
The technical data are shown in Figure
: System height (SH)
7 Preparation for use
7.1 Information on fabrication of a prosthesis
CAUTION
Incorrect alignment, assembly or adjustment
Injuries due to incorrectly installed or adjusted as well as damaged
prosthetic components
► Observe the alignment, assembly and adjustment instructions.
CAUTION
Initial use of the prosthesis by the patient
Fall due to lack of patient experience, incorrect alignment or incor
rect adjustment of the prosthesis
► For the safety of the patient, use a suitable device (e.g. parallel
bars, handrail, wheeled walker) during initial standing and walk
ing.
7.2 Bench alignment of the prosthesis
INFORMATION
The prosthetic components and prosthesis shown in the illustrations
are used as examples to illustrate the general process. The instruc
tions for use of the prosthetic components selected for the patient
contain detailed information and have to be used for fabrication of
the prosthesis.
► CAUTION! To enable the patient to stand safely, align the pros
thesis using the alignment images for the sagittal plane and front
3
al plane (Figure
on page 4) as well as the instructions for use
of all prosthetic components that were used.
2
on page 3.
21