8.0 TROUBLE
8.1
ERROR
MESSAGES
If an error message is shown, repeat the step that caused the message. If the error
message is still shown then contact your dealer for support.
ERROR
DESCRIPTION
CODE
Err 4
Initial Zero is greater
than allowed (4% of
maximum capacity
(+/‐2% for approved
units) when power is
turned on or when
the [‐>0/T<‐] key is
pressed.
Err 5
Keyboard Error.
Err 6
A/D count is not
correct when turning
the scale on.
Err 9
A/D count is not
stable when turning
the scale on.
Low Battery indicator. Internal battery may be
Unstable – balance
cannot get a stable
reading.
Red Light
Low Battery indicator. Internal battery may be
on front
panel
No Power when
turning on.
© Adam Equipment Company 2018
SHOOTING
POSSIBLE CAUSES
Weight on the pan
when turning on.
Excessive weight on the
pan when zeroing the
balance.
Improper calibration of
the balance.
Damaged load cell.
Damaged Electronics.
Improper operation of
the balance.
Load cell damaged.
Electronics damaged.
Load cell damage.
Something touching the
pan.
Air movement,
vibration or instability.
flat.
Possible damage to the
mechanics / Loadcell.
flat.
Internal battery may be
flat.
SOLUTIONS
Remove the transit screw make
sure the pan is fitted correctly.
Make sure the calibration weight is
in its off position.
Remove any weight from the
stainless steel pan.
Try recalibrating.
Turn the balance off and back on
again.
Remove any weight from the
stainless steel pan and turn the
balance off and back on again. Try
calibrating the balance.
There may be movement,
vibration or dirt on scale during
turning on. Make sure there is
nothing touching the pan. Make
sure the internal weight is in the
off position. Make sure the
balance is level. Try recalibrating.
Charge the battery or replace
internal rechargeable battery.
Make sure the balance is on a flat
surface and away from vibration.
Make sure the calibration weight is
in its off position.
Charge the battery or replace
internal rechargeable battery.
Charge the battery or replace
internal rechargeable battery.