EN
6.6 ADJUST
THE
)
MODELS
The internal mass value stored in memory can be adjusted to more closely match
the value of the user's external mass.
1) To set this parameter press the [Mode] key during self‐checking test when
turning on the balance.
2) After a few seconds, the display will show the first function F1 UNT.
3) Keep pressing the [Mode] key until F6 CA is displayed.
4) Press the [‐>0/T<‐] key to view the current settings.
5) The display will show the current value with the first digit flashing. To change
the value press [%] to shift the position, press [Mode] to increase the value
and press [Print] to decrease the value. The internal calibration weight should
only change between 95.000 and 105.000 grams or 495.00 and 505.00 grams
6) Press [‐>0/T<‐] to store a particular setting. The display will return
7) When CA is displayed, press the [Print] key to return to weighing or press
[Mode] to go to the next function.
NOTE: Changing the value of the internal weight will affect the calibration when
using the internal calibration. To check if your internal weight is correct you should
only use high quality weights that are accurate to the readability of the balance.
© Adam Equipment Company 2018
VALUE
OF
THE
INTERNAL
(depending upon the model).
to F6 CA
MASS
(N
OT POSSIBLE ON APPROVED
20