EN
6.2 SETTING
THE
The backlight and audible alert may be enabled or disabled by the user. If the
backlight is disabled, the battery life will be greater. The following settings are
available: (beep settings available in SW ver. 2.5 onwards)
bL AU
bL off
bL on
bP on
bP oFF
1) To set this parameter press the [Mode] key during self‐checking test when
turning on the balance.
2) After a few seconds, the display will show the first function F1 UNT.
3) Press the [Mode] key to select F2 EL .
4) Press the [‐>0/T<‐] key once to view the current settings for the backlight, and
twice to view settings for the alert beep.
5) To change the settings press the [Mode] key to scroll through other settings
as shown above.
6) Press [‐>0/T<‐] to store a particular setting. Press again until the display
returns to F2 EL
7) When F2 EL is displayed press the [Print] key to return to weighing or press
[Mode] to go to the next function.
6.3 SETTING
THE
1) To set this parameter press the [Mode] key during self‐checking test when
turning on the balance.
2) After a few seconds, the display will show the first function F1 UNT.
3) Keep pressing the [Mode] key until F3 SEr is displayed.
4) Press the [‐>0/T<‐] key to view the current settings.
© Adam Equipment Company 2018
BACKLIGHT
AND
Sets the backlight to operate automatically when a weight is
placed on the balance or a key is pressed.
Sets the backlight to be off.
Sets the backlight to be permanently on.
Sets audible alert beep on
Sets audible alert beep off
PRINTING
PARAMETERS
AUDIBLE
ALERT
BEEP
/
ACCUMULATION
16