Tabla de contenido
Idiomas disponibles

Idiomas disponibles

FORMULAS FOR CALCULATING RESULTS
Conversion of mass to volume
V = (w + e) x Z
Evaporation loss can be significant with low volumes. To determine mass loss, dispense
water to the weighing vessel, note the reading and start a stopwatch. See how much the
reading decreases during 30 seconds (e.g. 6mg = 0.2mg/s). Compare this to the pipetting
time from tareing to reading. Typically pipetting time might be 10 seconds and the mass
loss is 2mg (10s x 0.2mg/s) in this example. If an evaporation trap or lid on the vessel is
used the correction of evaporation is usually unnecessary.
The factor Z is for converting the weight of the water to volume at test temperature and
pressure. A typical value is 1.0032µl/mg at 22°C and 95 kPa. See the conversion table
on page 35.
Accuracy (systematic error)
Accuracy is the difference between the dispensed volume and the selected volume of a
pipette.
A = V -V
A = accuracy
0
V = mean volume
V = nominal volume
Accuracy can be expressed as a relative value: A% = 100% x A / V
Precision (random error)
Precision refers to the repeatability of the pipettings. It is expressed as standard deviation
(s) or coefficient of variation (cv)
n
S
(
V
V
-
S
=
i
i=1
n
-
When the Finnpipette Digital Multichannel is not in use, make sure it is stored in an upright
position. We recommend a Finnpipette stand for this purpose.
SHORT-TERM CHECKING
The pipette should be checked at the beginning of each day for dust and dirt on the outside
surfaces of the pipette.
Particular attention should be paid to the tip cone. No other solvents except 70 % ethanol
should be used to clean the pipette.
V = volume (µl)
w = weight (mg)
e = evaporation loss (mg)
Z = conversion factor for mg/µl conversion
s = standards deviation
v = mean volume
)
n = number of measurements
cv is the relative value of standard deviation.
cv = 100% x s / v

MAINTENANCE

8
0
Tabla de contenido
loading

Tabla de contenido