Malfunctions shown
Symbol ERR 1 is displayed –
Measurement error
ERR 2 – inflation error
ERR 3 – deflation error
ERR 330
No measuring or check displays
on the screen
Battery symbol
Faulty measurement values
Cuff is not pumped
Reasons
Movement of the arm or body during
the measurement. Do not talk during the
measurement
Pump problem
Deflation rate too high or too low, valve
problem
Air pressure exceeds 330 mmHg
Battery problem
No battery inserted – Battery incorrectly
inserted – Flat battery
Battery is flat
Moving your arm or body
during the measurement.
Too low or high position of wrist
Cuff does not fit
Cuff is not put on correctly
Heart rhythm malfunctions during measure-
ment
Battery is flat
Corrective Action
Repeat the measurement. Don't move the arm and
don't talk. In case of recurrence please contact our
after-sales service
Repeat measurement. In case of recurrence please
contact our after-sales service
In case of recurrence please contact our after-sales
service
Repeat measurement
Check batteries
Please check battery and replace if necessary
Please check battery and replace if necessary
Repeat the measurement. Don't move the arm and
don't talk.
Repeat measurement with correct posture
Check wrist circumference
Repeat measurement
Repeat measurement
Please check battery and replace if necessary.