Electromagnetic Compatibility Information
Table 1
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture's declaration - electromagnetic emissions
The KD-513LJ is intended for use in the electromagnetic environment specified below.
The customer or the user of the KD-513LJ should assure that it is used in such an environment.
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer's declaration – electromagnetic immunity
The KD-513LJ is intended for use in the electromagnetic environment specified below. The customer or the user of the KD-513LJ should
assure that it is used in such an environment.
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Power frequency (50/ 60 Hz)
magnetic field IEC 61000-4-8
Table 3
Compliance
Electromagnetic environment – guidance
Group 1
The KD-513LJ uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby elec-
tronic equipment.
Class B
The KD-513LJ is suitable for use in all establishments other than domestic and
those directly connected to the public low-voltage power supply network that
Not applicable
supplies buildings used for domestic purposes.
Not applicable
IEC 60601 test level
± 6 kV contact
± 8 kV air
3 A/m
Compliance level
± 6 kV contact
± 8 kV air
3 A/m
70
Electromagnetic environment – guidance
Floors should be wood, concrete or cera-
mic tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30 %.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a