Error Messages / Troubleshooting - innohaus ABM 47 Manual De Instrucciones

Tensiómetro para brazo
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

• The cuff integrity is maintained after 1,000 open–close
cycles of the closure.
• No component of this blood pressure monitor can be
maintained by the user. Circuit diagrams, component part
lists, descriptions, calibration instructions, or other
information, that will assist qualified service personnel to
repair parts that are designated as repairable, are avail-
able on request.
Disposal
Observe the local regulations for material disposal. Dispose
of the device in accordance with local regulations. If you
have any questions, please contact the local authorities re
sponsible for waste disposal.

9. Error messages / troubleshooting

In the event of errors, the error message
the display.
Error messages may appear if:
• systolic or diastolic pressure could not be measured (
or
appears on the display),
• systolic or diastolic pressure was outside the measure-
ment range ( or Lo appears on the display),
• the cuff is fastened too tightly or loosely (
pears on the display),
• the pump pressure is higher than 300 mmHg (
pears on the display),
• pumping up takes longer than 160 seconds (
on the display),
• there is a system or device error (
appears on the display),
• the batteries are almost empty
In such cases, repeat the measurement. Ensure that you
do not move or speak.
If necessary, reinsert or replace the batteries.
T echnical alarm – description
Should the recorded blood pressure (systolic or diastolic) lie
outside the limits specified in the section "Technical speci-
fications", the technical alarm will appear on the display
indicating either " " or "Lo". In such cases, you should seek
medical assistance and check the accuracy of your proce-
_ appears on
dure.
The limit values for the technical alarm are factory set and
cannot be adjusted or deactivated. These alarm limit val-
ues are accorded second priority under the standard IEC
60601-1-8.
The technical alarm is a non-locking alarm and must not be
reset. The signal shown on the display will disappear auto-
or
ap-
matically after about 8 seconds.
19
ap-
appears
,
,
or
.