VIII.-OTHER IMPORTANT MESSAGES THAT APPEAR IN THE SCALE / OTROS
MENSAJES IMPORTANTES QUE APARECEN EN LA BÁSCULA
This message indicates that a weight that exceeds the maximum capacity has been placed on the
scale. To correct this, simply remove the over weight.
This message appears in two different situations:
A) when the scale is turned on with an excessive weight on the plate. To correct this situation
remove the weight that is on the scale, turn off the scale and then turn it on again.
B) when the scale was handled roughly and/or got hitted during the shipment, which caused
irreversible damages in this operation. In this case call your authorized distributor.
This message indicates a negative weight, this could happen if you lift or remove the scale's plate.
To correct this put the plate back in its place and press the
key or turn the scale off and then
turn it on again.