The test does
Blood sample is too small
not start after
applying the
blood sample.
Sample applied after meter
turns off
Defective meter or test strip
What It Means
What To Do
1. See test strip instructions for use for
re-application instructions.
2. If the countdown still does not start, remove
the used strip and discard it properly.
3. Repeat the test using a new test strip.
4. If the test still does not start, contact
Customer Service.
1. Remove the used strip and discard it
properly.
2. Review the testing instructions.
3. Repeat the test using a new test strip.
4. If the test still does not start, contact
Customer Service.
1. Remove the used strip and discard it
properly.
2. Repeat the test using a new test strip.
3. If the test still does not start, contact
Customer Service.
38