Dentsply Maillefer CAVITY ACCESS KIT Instrucciones De Uso página 5

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 24
Step-by-step procedure
A. Devices
Operation
Operating mode
1.
Disassembling
- Disassemble the
2.
Pre-Disinfection
- Soak all instruments
3.
Rinsing
- Abundant rinsing (at
4a. Automated Cleaning
- Place the devices
with washer-disinfector
- Put them in the
OR
4b. Manual Cleaning
- Place the devices
or assisted by an
ultrasonic device
- Immerse in the
5.
Rinsing
- Abundant rinsing (at
6.
Inspection
- Inspect devices and
- Assemble the
7.
Packaging
- Place the devices
DFU CAVITY ACCESS KIT - F190342.X/ 06 / 2005 updated 05/2010
a3. contra angle
a2. Implantable devices
a1. Instruments
Warning
Silicone stops have to be removed.
device, if required
- Follow instructions and observe concentrations and immersion
immediately after
times given by the manufacturer (an excessive concentration
use in a detergent
may cause corrosion or others defects on instruments).
and disinfecting
- The disinfecting solution should be aldehyde free (to avoid
solution combined
blood impurities fixation) and without di- or triethanolamines as
with proteolytic
corrosion inhibitor.
enzyme if possible.
- Do not use disinfecting solutions containing Phenol or any
products which are not compatible with the instruments (See
general recommendations).
- For visible impurities observed on instruments a pre-cleaning is
recommended by brushing them manually with soft material.
- Use quality water in accordance with local regulations.
least 1 min)
- If a pre-disinfectant solution contains a corrosion inhibitor, it is
recommended to rinse the instruments just before the cleaning.
- Discard any instruments with large obvious defects (broken,
in a kit, support or
bent).
container to avoid
- Avoid any contact between instruments or posts when placing
any contact between
in the washer-disinfector use kits, supports or container.
instruments or posts.
- Follow instructions and observe concentrations given by the
manufacturer (see also general recommendations).
washer-disinfector
- Use only approved washer-disinfector according to EN ISO
(Ao value > 3000
15883, maintain and calibrate it regularly.
or, at least 5 min at
90 °C).
- No visible impurities should be observed on the instruments.
in a kit, support or
- Discard any instruments with large obvious defects (broken,
container to avoid
bent, and twisted).
any contact between
- Follow instructions and observe concentrations and time given
instruments.
by the manufacturer (see also general recommendations).
- The disinfecting solution should be aldehyde free and without
disinfecting solution
di- or triethanolamines as corrosion inhibitor.
with cleaning
properties, assisted
by an ultrasonic
device if suitable.
- Use quality water in accordance with local regulations.
least 1 min)
- If a disinfecting solution contains a corrosion inhibitor, it
is recommended to rinse the instruments just before the
autoclaving.
- Dry on a single use non-weaved cloth, or with a drying machine
or filtered compressed air.
- Dirty instruments must be cleaned and disinfected again.
sort out those with
- Discard instruments which show any deformations (bent,
defects.
twisted), damages (broken, corroded) or defects (loss of colour
coding or marking) affecting the resistance, the safety or the
devices (stops)
performance of the instrument or posts.
- Protect carbon steel bur with corrosion inhibitor before
packaging.
- For Contra Angle: lubricate the device with an adequate spray
before packaging
- Avoid any contact between instruments or posts during
in a kit, support or
sterilization. Use kits, supports or containers.
container to avoid
- Check the validity period of the pouch given by the
any contact between
manufacturer to determine the shelf life.
instruments or posts
- Use packaging which are resistant up to a temperature of
and pack the devices
141°C (286°F) and in accordance with EN ISO 11607.
in "Sterilisation
pouches".
DENTSPLY MAILLEFER
1338 Ballaigues
Switzerland
Following uses
First use
X X X X
X X X
X X X
X X X X
X
X
X
X
X X X
5 / 72