2. Check that the system's power cord plug is fully seated
in an approved AC wall outlet.
3. Check that the wall outlet is functional.
Symptom:
System will not operate: Power ON Indicator is
illuminated
1. If the office has more than one foot control, test each
to ensure that the proper foot control is being used.
With a handpiece and insert installed, depress the
foot control to the first position. The system should
dispense water. If none of the foot controls operate
the system, continue to the next step.
2. Resynchronize one foot control to the system (see
Section 6.8 Foot Control Synchronization).
Symptom:
System operates: No water flow to insert tip or
handpiece overheats
1. Assure that handpiece lavage control is properly
adjusted.
2. Check for clogged insert.
3. Check that dental office water supply valves are open.
4. If the system is connected to DualSelect Dispensing
System, check that fluid level in the selected bottle is
sufficient. Make sure valves are open when using
external water source.
5. Check that the water line filter is clean. Replace filter
if needed.
Symptom:
System operates: No insert cavitation
1. Check that the Power Level Control is not in Rinse
Mode.
2. Check the insert for damage and that it is properly
installed in the handpiece.
3. Check that the handpiece is properly installed to the
cable assembly.
4. If Steri-Mate grip is used on the handpiece, verify that
the grip is flush with the hard plastic of the insert port.
5. Turn the system's Main Power Switch to the OFF (0)
position. Wait 5 seconds and turn the system back ON.
6. If problem still exists, replace both "AA" batteries in
foot control with new "AA" batteries (Refer to Section
6.7) or connect auxiliary foot control cable.
Symptom:
System operates: Purge Mode will not function
– icon flashing
1. Check that there is no insert in the handpiece.
2. Check that handpiece is properly installed to the cable
assembly.
Symptom:
System operates: Service Indicator blinking
•
Slow Blinking (1 blink per second)
The system is not operating within factory
specifications.
1. Remove insert.
2. Turn Main Power Switch OFF, (O) position. Wait five
seconds. Turn unit ON, (I) position.
3. Operate Purge function.
4. If service indicator still blinks, refer to Section 10.2
Technical Support and Repairs to have unit
serviced as soon as possible.
•
Fast Blinking (3 blinks per second)
– Indicates improper set-up
1. If insert is in the handpiece, remove. Verify the
handpiece is properly seated and depress the foot
control for 2 seconds. If blinking stops, the system
is ready for use. If blinking remains, continue to the
next step.
2. Attach a NEW handpiece and depress foot control
for 2 seconds. If blinking stops, the system is
ready for use. Discard the old handpiece or return if
within warranty. If blinking remains, continue to the
next step.
3. Install and fully seat an insert into handpiece.
Depress foot control for 2 seconds. If blinking
stops, unit is ready for use. If blinking remains,
continue to the next step.
4. Install and fully seat a NEW insert in handpiece and
depress foot control for 2 seconds. If blinking
stops, system is ready for use. Discard old insert
or return if within warranty. If blinking remains,
refer to Section 10.2 Technical Support and
Repairs to have unit serviced as soon as possible.
Symptom:
System operates: Service Indicator illuminated
1. Ensure that the base unit has adequate ventilation and
is not near a heat source (i.e. radiator, heat lamp,
sunlight or other heat producing operatory equipment).
2. Turn Main Power Switch to the OFF (O) position. Allow
system to cool for 10 minutes and turn system ON, (I)
position. Verify light is not illuminated.
3. If light is still illuminated, refer to Section 10.2
Technical Support and Repairs to have unit serviced as
soon as possible.
10.2 Technical Support and repairs
For technical support and repair assistance call DENTSPLY
Professional Cavitron Care
800-989-8826 Monday through Friday, 8:00 A.M. to 5:00
P.M. (Eastern Time). For areas outside the U.S., contact your
local DENTSPLY Professional representative.
14