EN-6
4. MENU SETTINGS
Press and hold SET until [SEt-Fx] is displayed, where x:
0 = Set checkweigh limits and mode
1 = Set user settings
2 = Restore user settings to factory defaults
3 = Set tare capacity range
4.1 Checkweigh Over and Under Limits
•
Press and hold SET until [SEt-F0] is displayed.
•
Press YES to view the upper limit [000000] (or last value
will illuminate.
•
Press
,
,
•
Press YES to confirm, then view the lower limit [000000] (or last value saved).
The Under LED will illuminate.
•
Press
,
,
divisions, and less than upper
•
Press YES to confirm, then view the alarm setting [rUn x], where x:
0 = Checkweigh mode is disabled
1 = Checkweigh LEDs are active, beeper will sound when result is outside the
Accept range
2 = Checkweigh LEDs are active, beeper will sound when result is within the
Accept range
3 = Checkweigh LEDs are active, beeper is disabled
•
Press
to step through available settings, then press YES to select.
•
Press SET to save and exit.
Note: The measuring unit of the limit values will follow the active unit (see Section
4.2).
to edit the upper limit value (must be less than scale capacity).
to edit the lower limit value
limit).
OHAUS Valor
saved).
(must be minimum of 20d scale
TM
1000W Series
The Over LED