Invacare recommends using the XPO
4 ft (1.2 m) cannula. The tubing must be connected to the
XPO
concentrator oxygen outlet as shown below.
2
Replace the nasal cannula on a regular basis. Check with
your equipment provider or physician to determine how
often the cannula should be replaced.
DO NOT share cannulas between patients.
DETAIL "A"
DETAIL "B"
1148112-F~12
concentrator with a
2
1.
Connect the nasal cannula to the XPO
outlet port A (Detail "A").
2.
Place the cannula B over your ears and position the prongs in
your nose as instructed by your health care provider or cannula
manufacturer (Detail "B").
5.3 Using the XPO
2
1.
Turn the XPO
concentrator On by pressing the On/Off button.
2
2.
Breathe normally through your nose. Breathing through your
mouth may result in less effective oxygen therapy. The XPO
concentrator will put out a measured pulse of oxygen each
time it detects inhalation. The blue flow setting indicator light
will blink each time an inhalation is detected and the XPO
concentrator outputs a pulse of oxygen.
3.
Turn the XPO
concentrator Off by pressing and holding the
2
power On/Off button for approximately two seconds.
The time required for the XPO
maximum oxygen concentration after being turned On is
approximately 30 minutes.

Recommendations for Optimal Performance

Operating Temperature:
Relative Humidity:
Transport/Storage
Temperature:
Electrical:
concentrator oxygen
2
Portable Concentrator
concentrator to reach
2
41° F to 95° F (5° C to 35° C)
15% to 60%
-2° F to 140° F (-20°C to 60°C)
Allow unit to warm, or cool,
to operating temperature range
before using.
No extension cords.