• Discard the set after each use. Do not clean or re-sterilize.
• Reuse of the infusion set may cause infection, site irritation, or damage
to the cannula/needle. A damaged cannula/needle may lead to inaccurate
medication delivery.
• The Accu-Chek Tender is for subcutaneous use only.
• Always tighten the luer-lock firmly into the pump adapter. If this is not done,
danger of leakage can result, due to a loose connection between the infusion
set and the pump. Avoid using any tools to screw the luer-lock connector into
place. Otherwise damage could result to the luer-lock connector.
• Do not leave air bubbles in the Accu-Chek Tender. Prime thoroughly.
• A small percentage of Accu-Chek Tender using soft cannulas can accidental-
ly get crimped during insertion or displaced while in use. If this should occur
you should immediately use a new Accu-Chek Tender.
• Check frequently to make sure the soft cannula remains firmly in place.
Because the cannula is so soft, you will not notice any pain if it is pulled out.
The soft cannula must always be completely inserted in order to receive the