Microphone Calibration - Phonic PAA6 Manual Del Usuario

Ocultar thumbs Ver también para PAA6:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

MICROPHONE CALIBRATION

As the PAA6 comes already calibrated, you should never actually need to calibrate your unit. If measure-
ment data or the operation of the unit begins to seem abnormal, however, it may be necessary to perform
a calibration. Anyone can calibrate the PAA6 and regain accurate sound pressure level measurement,
provided they have a quality sound level calibrator with 1/2" diameter adapter that sends out a 1 kHz
tone. A B&K TYPE 4231 sound level calibrator is suggested.
Procedure:
1. Enter the System > Calibration > Mic menu (figure 69).
2. As you probably cannot calibrate both microphones concurrently, it's advised that you turn one
microphone off while you calibrate the other. Push the MIC1 or MIC2 icons at the top right-hand
side of the screen to turn one of the microphones off.
3. Place the PAA6's microphone within a sound level calibrator with a microphone connector of a 1/2"
diameter.
50
figure 69
PAA6

Capítulos

Tabla de contenido
loading

Tabla de contenido