Writing single register: STOP rotation
Request
Field
Address
Control command
High start address
Low start address
High register value
Low register value
High CRC
Low CRC
Virtual serial port
Baudrate
Bits
Stop Bit
Patiry
Maximum number of registers
for single request
5. Menu
Hold the control knob (F) for 3 seconds to enter into the setting mode when stirring is off.
Turning the control knob (F) it is possible to pass from one of the following parameters to the next one. Click the knob to modify:
Parameter shown
on display
NOdE
OHS 200
OHS 100
LINt
OHS 60
OHS 40
OHS 20
COun
UPGr
rSEt
Click the stirring control knob (F) to display the set value. Turn the knob to set the new value and click to confirm.
To leave settings mode wait 10 seconds without touching any knob or click the lock key.
(Hex)
0x64
0x06
0x00
0x70
0x00
0x01
0x24
0x40
9600
8
1
None
24
Default value
StOP
400 (I)
2000 (II)
1300
2000
2000
2000
nO
Range
Set starting mode:
Stop: when the instrument is switched on through
the main switch, OFF is on display
StOP - run
Run: when the instrument is switched on, it
restarts to work with the last set points set
100 – 400
100 - 2000
100 – 1300
Set full scale speed:
It limits the maximum value of speed set point
100 – 2000
(from 100rpm, step 100rpm)
100 - 2000
100 - 2000
Running time. Days (24H):
Motor working time is displayed in hours until
9999 hours (around 416 days).
After, it's shown in days starting from 417 and a
dot turns on as following, to differentiate days
from hours ".417"
Software upgrading:
It allows to install a new software version
Reset:
YES - nO
It allows to restore default setting parameters
8
Description
Stop rotation
Description