Recommended separation distances between portable and mobile RF
The
d
evice
is intended for use in an electromagnetic environment in which ra- diated
RF disturbances are controlled. The customer or the user can help prevent
electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the
AirForce Mini
communications equipment.
Rated maxi-
mum output
power of
transmitter W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
6 3
communications equipment and the
as recommended below, according to the maximum output power of the
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
(out ISM and amateur
radio bands)
d=1.2√p
0.12
0.38
1.2
3.8
12
150 kHz to 80 MHz
(in ISM and amateur
radio bands)
d=0.6√p
0.06
0.19
0.6
1.9
6
AirForce Mini
80MHz to
800MHz to
800MHz
d=1.2√p
0.12
0.38
1.2
3.8
12
2.7GHz
d=2.3√p
0.23
0.73
2.3
7.3
23