Troubleshooting - Gima THP59J Manual De Uso Y Mantenimiento

Termómetro infrarrojo de oído profesional
Ocultar thumbs Ver también para THP59J:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Trouble shooting
Error Message
ENGLISH
Problem
The probe cover has not been
installed well. will flash
LCD screen and can't take the ear
temperature
Error 5~9, the system is not
functioning properly.
Measurement before device
stabilization.
The device showing a rapid
ambient temperature change.
The ambient temperature is not
within the range between 10 °C
and 40 °C (50 °F ~104 °F).
Temperature taken is higher than
+42.2 °C (108°F)
Temperature taken is lower than
+34 °C (93.2 °F)
Device cannot be powered on to
the ready stage.
Solution
Please check the setting of the
probe cover again.
on the
(Refer to the section of "Place
the Probe Cover" )
Unload the battery, wait for 1
minute and repower it. If the
message reappears, contact the
retailer for service.
Wait until all the icons stops
flashing.
Allow the thermometer to rest in
a room for at least 30 minutes at
room temperature: 10 °C and 40
°C (50 °F ~104 °F).
Check the integrity of the
probe cover and take a new
temperature measurement.
Make sure the probe cover
i s c l e a n a n d t a k e a n e w
temperature measurement.
Change with a new battery.
12
Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

25575

Tabla de contenido