Beckman Coulter HHPC 6+ Manual Del Usuario página 5

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 108
Electromagnetic Compatibility (EMC)
This device complies with the emissions and immunity requirements as specified in the EN/IEC
61326 series of Product Family Standards for a "basic electromagnetic environment." Such
equipment is supplied directly at low voltage from public mains network. This equipment is not
intended for residential use.
CAUTION
This device generates, uses, and can radiate unintentional radio-frequency (RF)
energy. If this device is not installed and operated correctly, this RF energy can
cause interference with other equipment. It is the responsibility of the end user to
be sure that a compatible electromagnetic environment for the device can be
maintained so that the device operates as intended.
This equipment is designed for use in a PROFESSIONAL FACILITY ENVIRONMENT. It is likely to
perform incorrectly if used in a HOME ENVIRONMENT. If it is suspected that performance is
affected by electromagnetic interference, correct operation may be restored by increasing the
distance between the equipment and the source of the interference.
In addition, other equipment can radiate RF energy to which this device is sensitive. If one suspects
interference between this device and other equipment, Beckman Coulter recommends the
following actions to correct the interference:
1
Evaluate the electromagnetic environment before installation and operation of this device.
2
Do not operate this device close to sources of strong electromagnetic radiation (for example:
unshielded intentional RF sources), as these can interfere with proper operation. Examples of
unshielded intentional radiators are handheld radio transmitters, cordless phones, and cellular
phones.
3
Do not place this device near medical electrical equipment that can be susceptible to
malfunctions caused by close-proximity to electromagnetic fields.
4
This device has been designed and tested to CISPR 11, Class A emission limits. In a domestic
environment, this device may cause radio interference, in which case, you may need to take
measures to mitigate the interference.
Class 1 laser product
This instrument is classified as a Class 1 laser product. This product complies with IEC/EN
60825-1:2014 and 21 CFR 1040.10 except for deviations pursuant to Laser Notice No. 56, dated
May 8, 2019
DOC026.97.80271, Edition 9
HHPC 6+, HHPC 3+, HHPC 2+ USER MANUAL
General information
5 / 26
loading

Productos relacionados para Beckman Coulter HHPC 6+

Este manual también es adecuado para:

Hhpc 3+ 0.3 umHhpc 3+ 0.5 umHhpc 2+