Set the data export mode
There are two file formats for data export. The user selects the format in the Communication Setup
screen.
If an administrator password is set in the security screen, the data export mode cannot be selected
by lower level users. This gives a way to make sure that the secure .xls mode is used when necessary.
1
Go to the Communication Setup menu.
2
Click the radio button for one of the Data Export modes.
Table 3
Option
.tsv
.xls
Move data to a USB memory stick
1
Connect the memory stick to the USB port on the bottom of the instrument.
2
Navigate to the Buffered Data Icon and push the
3
In the Buffered Data screen, push the
a green arrow is highlighted. Push the
The data export starts automatically. The display shows a successful export message when the
export is complete. Do not remove the memory stick before this message shows. For buffers
with a large amount of data, the export may take a few seconds.
4
Push the
SELECT
5
Remove the USB stick from the instrument and connect the stick to the USB port on the
computer.
6
In Windows Explorer, navigate to the computer drive for the USB memory stick.
DOC026.97.80271, Edition 9
Description
This format gives an Excel compatible tab separated variable (.tsv) file. The file is not locked
for editing and is not CFR 21 Part 11 secure. This format supports a maximum of 10,000 data
records.
This format gives an Excel (.xls) spreadsheet file. The file is locked for editing and is CFR 21
Part 11 secure. This format supports a maximum of 5,000 data records.
button to remove the message.
HHPC 6+, HHPC 3+, HHPC 2+ USER MANUAL
button.
SELECT
until the icon that shows the USB stick and
LEFT ARROW
button.
SELECT
Operation