LIMIT DCM 3 Instrucciones De Uso página 11

Idiomas disponibles
  • ES

Idiomas disponibles

2.2 AC voltage/frequency
Range
4.000 V
40.00 V
400.0 V
600 V
Voltage
frequency
10 Hz~60 kHz
Caution:
Short press "SELECT" in AC voltage/Hz scale to enter the Hz function;
The input impedance is about 10 MΩ
Current/voltage frequency response: 45 Hz ~ 400 Hz, displays true RMS
value
Accuracy guarantee range: 1%~100%
AC crest factor of non-sinusoidal wave can reach 3.0 at 4000 counts, while
it can only reach 1.8 at 6000 counts, the additional error should be added
for the corresponding crest factor as follows:
a. Add 3% when the peak factor is 1 ~ 2
b. Add 5% when the peak factor is 2 ~ 2.5
c. Add 7% when the peak factor is 2.5 ~ 3
3. Continuity/Diode measurement
Function
Range
400.0 Ω/
600.0 Ω
4.000 V/
6.000 V
4. Resistance
Range
400.0 Ω
4.000 kΩ
40.00 kΩ
400.0 kΩ
4.000 MΩ
40.00 MΩ
Resolution
0.001 V
0.01 V
0.1 V
1 V
0.01 Hz~0.01 kHz
Resolution
≤30 Ω: Consecutive
beeps
0.1 Ω
≥31 Ω: No beep
The median: uncertain
The open circuit voltage
is about 4 V
For the silicon PN
0. 001 V
junction diode, the
voltage value is generally
about 0.5~0.8 V.
Resolution
0.1 Ω
0.001 kΩ
0.01 kΩ
0.1 kΩ
0.001 MΩ
0.01 MΩ
Accuracy
± (1.0%+5)
± (0.8%+5)
± (0.5%+2)
Accuracy
Accuracy
± (1.0%+2)
± (0.8%+2)
± (2.5%+5)
Overload
protection
600 Vrms
Overload
protection
600 Vrms
Overload
protection
600 Vrms
9
GB
loading