ProMed PBM-4.2 Instrucciones De Empleo página 139

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 60
EMC DECLARATION
Voltage dips, short interruptions
and voltage variations on power
supply input lines
IEC 61000-4-11
Power frequency magnetic field
IEC 61000-4-8
Conduced RF
IEC61000-4-6
Radiated RF
IEC61000-4-3
NOTE UT is the a.c. mains voltage prior to application of the test level.
Table 3
Guidance and manufacturer's declaration - electromagnetic Immunity
Radiated RF
Test
IEC61000-4-3
Frequency
(MHz)
(Test speci-
fications for
385
ENCLOSURE
PORT
IMMUNITY
450
to RF wireless
commu-
nications
equipment)
710
745
780
810
870
930
1720
1845
1970
2450
5240
5500
5785
0 % 0.5 cycle
At 0º, 45 º, 90 º, 135 º, 180 º,
225 º, 270 º and 315 º
0 % 1 cycle
And 70% 25/30 cycles
Single phase: at 0
0 % 300 cycle
30 A/m
50Hz/60Hz
150KHz to 80MHz:
3Vrms
6Vrms (in ISM and amateur
radio bands)
80 % Am at 1kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Band
Service
(MHz)
380 –
TETRA 400
390
430 –
GMRS 460,
470
FRS 460
704 –
LTE Band 13,
787
17
800 –
GSM
960
800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
1700 –
GSM 1800;
1990
CDMA
1900;
GSM 1900;
DECT;
2400 –
Bluetooth,
2570
WLAN,
802.11
b/g/n,
RFID 2450,
LTE Band 7
5100 –
WLAN
5800
802.11
a/n
139
Not application
30 A/m
50Hz/60Hz
Not application
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Modulation Modulation
(W)
Pulse
1,8
modulation
18 Hz
FM
2
± 5 kHz
deviation
1 kHz sine
Pulse
0,2
modulation
217 Hz
Pulse
2
modulation
18 Hz
Pulse
2
modulation
217 Hz
Pulse
2
modulation
217 Hz
Pulse
0.2
modulation
217 Hz
Distance
IMMUNITY
(m)
TEST LEVEL
(V/m)
0.3
27
0.3
28
0.3
9
0.3
28
0.3
28
0.3
28
0.3
9
DE
EN
FR
IT
ES
NL
RU
PL
SE
AR
loading

Este manual también es adecuado para:

U80eh