Error
Description Potential cause and remedy
«ERR 3»
No pressure
An adequate pressure cannot be gener-
in the cuff
ated in the cuff. A leak may have
AQ
occurred. Check that the cuff is correctly
connected and is not too loose. Replace
the batteries if necessary. Repeat the
measurement.
«ERR 5» Abnormal
The measuring signals are inaccurate
result
and no result can therefore be displayed.
Read through the checklist for
performing reliable measurements and
then repeat the measurement.*
«HI»
Pulse or cuff
The pressure in the cuff is too high (over
pressure too
299mmHg) OR the pulse is too high
high
(over 199 beats per minute). Relax for 5
minutes and repeat the measurement.*
«LO»
Pulse too
The pulse is too low (less than 40 beats
low
per minute). Repeat the measurement.*
* Please immediately consult your doctor, if this or any other problem
occurs repeatedly.
If you think the results are unusual, please read through the
information in «Section 1.» carefully.
24
8. Safety, care, accuracy test and disposal
m
Safety and protection
 Follow instructions for use. This document provides important
product operation and safety information regarding this device.
Please read this document thoroughly before using the device and
keep for future reference.
 This device may only be used for the purposes described in these
instructions. The manufacturer cannot be held liable for damage
caused by incorrect application.
 This device comprises sensitive components and must be treated
with caution. Observe the storage and operating conditions
described in the «Technical specifications» section.
 The cuffs are sensitive and must be handled with care.
 Only pump up the cuff once fitted.
 Do not use this device if you think it is damaged or notice anything
unusual.
 Never open this device.
 Read the additional safety information provided within the indi-
vidual sections of this instruction manual.
 The measurement results given by this device is not a diagnosis.
It is not replacing the need for the consultation of a physician,
especially if not matching the patient's symptoms. Do not rely on