7.1
Changing storage batteries
To replace the nickel-metal hydride (NiMH) storage battery, use a Phillips
2
screwdriver to unscrew the Phillips screw (1) on the bottom side of the SIRO-
Endo Pocket and remove the cover (2) of the storage battery compartment.
Lift out the storage battery block (3), pull the plug (4) of the connecting lead
3
and remove the defective storage battery block.
Insert the new storage battery block by following the same procedure in
reverse order.
Before closing the storage battery compartment, make sure that the addi-
tional weight (A) is properly inserted.