5. Perform the usual tests for
tolerance testing of the pressure reading.
leak testing of the pressure circuit (take
account of the cuff's settling time of one minute).
6. Switch off the unit and restore the original connec-
tions.
C) Security
For security purposes the two halves of the unit can be
joined together with a security mark. Alternatively the
hole for the casing connector screw on the rear side of
the unit can be secured by sealing it.