Display Indications / Error Messages - BEHA TELARIS 0100 Instrucciones Para El Manjeo

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Display Indications

8.0 Display Indications / Error Messages

Measurement Range Message
UAC
>440V
50Hz
UAC
25V
---
UDC
>440V
440V DC
UDC
>-440V
500V
RINS
>199,9MΩ
500V
ZL, IPSC
>1999Ω
LOW-Ω
>19,99Ω
LOW-Ω
COMP
RINS, LOW-W
RCD, ZL, IPSC
PE Contact Electrode
is touched
RCD
RCD
>500ms
5V UL:50V
RCD
200ms
5V UL:50V
RCD
5V UL:50V
38
Uext
,
Limit
Limit
Limit
345ms
Cause
overflow, voltage present exceeds
440V AC
frequency of AC voltage outside
range 10.0 ... 99,9Hz
overflow, voltage present exceeds
overflow, voltage present falls
below -440V DC
insulation resistance exceeds or is
greater than 199,99MΩ
loop impedance higher
than 1999Ω
Low ohm resistance higher
than 19.99Ω
Test lead resistance has been
properly compensated.
External voltage present, measure-
ment may not be carried out.
PE Error
Measurement of trip current, or
test of time delayed RCD.
Trip time exceeds 500ms or RCD
has not tripped.
Trip time exceeds 200ms or RCD
has not tripped.
RCD has tripped at 345 ms. As trip
time exceeds 200ms, the RCD is
not in order!
Tabla de contenido
loading

Tabla de contenido