Warnings
English
The UV lamp contains mercury and may be considered
hazardous material in your local area. Dispose of these
items in accordance with federal, state, or local government
regulations.
In the case of any broken or damaged UV lamp, handle the
remains in accordance with your organization's toxic waste
handling procedure and dispose of them in accordance with
federal, state, or local government regulations.
The WEEE Directive 2012/19/EU applies to batteries and
requires their removal and separate collection. Once removed
the used batteries are governed by the Battery Directive
2006/66/EC and its amendments. Dispose of the used
batteries according to the manufacturer's instructions and local
disposal/collection requirements.
Operation of the Analyzer without the in-line filter on the sample
inlet line will damage the Analyzer and void the warranty. To
avoid damaging the Analyzer, install the filter and replace the
filter element, as needed.
Make sure the DI water reservoir is full, particularly when
running samples with high TOC or high salt concentrations.
After running high TOC or salt samples, always rinse the
Analyzer by running low-TOC DI water through a sample cycle.
Refer to the Analyzer's operation and maintenance manual
for instructions on filling the DI water reservoir.
To avoid damage to the DI water pump, ensure that the DI
water reservoir is filled before turning on the Analyzer.
To avoid false TOC readings and possible damage to the
Analyzer, always make sure the sample inlet is open and the
DI water reservoir is filled before turning on the Analyzer.
Always close the door of the iOS system when not sampling
from a vial.
To protect the Analyzer from damage, always drain the water
from the Analyzer prior to shipping.
When cleaning the Analyzer, always apply liquid to the cloth
directly, and do not spray liquids directly on the Analyzer.
Do not use water or cleaners on the touchscreen, as water can
damage or discolor the polarizer. Refer to the next section for
recommendations on cleaning the touchscreen.
Wipe away any condensation immediately.
Operating the Analyzer in Online mode with no sample
flow and with the Flow Sensor off can result in erroneous
measurements.
Standards for the Sievers M500 Online TOC Analyzer are
acidified. If preparing your own standards and using glass
vials, ensure to acidify the standards to a pH of 3, or the
system protocols will yield poor results.
Ensure to center the vent tube within the tubing fittings to
prevent pinching beneath the vent tube and prevent the
sample from flowing.
To avoid injury and to ensure accurate analysis results, ensure
that the Analyzer door is closed and latched during normal
operation.