Immunity to proximity fields from RF wireless communication
equipment
Test standard
EN 61000-4-3
Voltage dips
Test standard
EN 61000-4-11
Voltage interruptions
Test standard
EN 61000-4-11
Conducted / Radiated RF emission
Test standard
EN 55011
Harmonic current emissions
Test standard
IEC 61000-3-2
Voltage changes, voltage fluctuations and flicker
Test standard
EN 61000-3-3
Immunity test level
385 MHz, 27 V/m, 18 Hz PM (50 % duty cycle square wave)
450 MHz, 28 V/m, FM +/- 5 kHz deviation, 1 kHz sine
710 MHz, 9 V/m, 217 Hz PM (50 % duty cycle square wave)
745 MHz, 9 V/m, 217 Hz PM (50 % duty cycle square wave)
780 MHz, 9 V/m, 217 Hz PM (50 % duty cycle square wave)
810 MHz, 28 V/m, 18 Hz PM (50 % duty cycle square wave)
870 MHz, 28 V/m, 18 Hz PM (50 % duty cycle square wave)
930 MHz, 28 V/m, 18 Hz PM (50 % duty cycle square wave)
1720 MHz, 28 V/m, 217 Hz PM (50 % duty cycle square
wave)
1845 MHz, 28 V/m, 217 Hz PM (50 % duty cycle square
wave) 1970 MHz, 28 V/m, 217 Hz PM (50 % duty cycle
square wave) 2450 MHz, 28 V/m, 217 Hz PM (50 % duty
cycle square wave) 5240 MHz, 9 V/m, 217 Hz PM (50 % duty
cycle square wave) 5500 MHz, 9 V/m, 217 Hz PM (50 % duty
cycle square wave) 5785 MHz, 9 V/m, 217 Hz PM (50 % duty
cycle square wave)
Immunity test level
0 % U_T; 0.5 cycle at 0°, 45°, 90°, 135°, 180°, 225°, 270° and
315°, 0 % U_T; 1 cycle at 0°, 70 % U_T; 25 cycles at 0°
Single phase at 0°
Immunity test level
0 % U_T; 300 cycle
Emission test class and group
Class B, group 1
Emission test class and group
Class A
Emission test class and group
Tested according to the standard.
en
23