Pressalit MCT 3 R9612717 Instrucciones De Uso Y Mantenimiento página 22

Camilla para cambio
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 148
Electromagnetic compatibility
• Use of accessories, transducers and cables other than those specified or
provided by the manufacturer could result in increased electromagnetic
emissions or decreased electromagnetic immunity of this equipment
and result in improper function.
• Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than
30 cm (12 inches) to any part of the unit, including its cables, to avoid
any degradation of the performance of this unit.
• Use of this equipment adjacent to or stacked with other equipment
should be voided because it could result in improper operation. If such
use is necessary, this equipment and the other equipment should be
observed to verify that they are operating normally.
Electromagnetic immunity / emission
Electrostatic discharge
en
Test standard
EN 61000-4-2
Radiated RF electromagnetic field
Test standard
EN 61000-4-3
Electrical fast transient/burst
Test standard
EN 61000-4-4
Surge immunity
Test standard
EN 61000-4-5
Immunity to conducted disturbances induced by RF fields
Test standard
EN 61000-4-6
22
WARNING
!
Immunity test level
+/- 8 kV contact, +/- 2, 4, 8, 15 kV air
Immunity test level
15 V/m 80 MHz to 1 GHz, 15 V/m 1,01 GHz to 6 GHz
80 % AM 1 kHz
Immunity test level
+/- 1 kV , +/- 2 kV, 100 kHz repetition frequency
Immunity test level
Line-to-line: +/- 0.5 kV, +/-1 kV
Line-to-ground: +/- 0.5 kV, +/-1 kV, +/- 2 kV
Immunity test level
3 V (6 V in ISM bands and amateur radio bands)
0,15 to 80 MHz , 80 % AM at 1 kHz
loading