AC/DC Voltage Measurement
Connected loads generating EMV disturbance or harmonic waves may impair the display
quality.
1)
Select the "UAC/DC" measurement range using the measurement function selection
dial (19).
2)
Connect the red test lead to the "UAC/DC" socket and the black test lead to the
"COM" socket.
3)
Connect the test leads to the UUT.
4)
Select between AC or DC voltage pressing the "AC/DC" key (17).
• When switching on or changing to the "UAC/DC" measurement function, the default
setting is always "AC".
5)
Read the measurement result on the display screen.
Overvoltage Protection for Voltage Measurement
If a external voltage of maximum 600 V AC/DC is connected by mistake to the UAC/DC
measurement socket, ISO-PRO will not be damaged. The instrument displays ">600V". After
removal of the overvoltage wait for approx. 20 s to obtain correct measurement values.
• If an AC voltage >25 V is measured within the DC voltage range, the instrument displays
"Uext".
• This is also valid for the AC range if measuring a DC voltage > 25 V.
19