Maintenance
Expected service life of the shower table is 10 years, provided that
maintenance is carried out as described below:
It is recommended on a weekly basis to run the motors on the product to their
maximum positions by raising the shower table to its full height, and then
lowering it to its lowest position, so that the internal parts are lubricated .
It is recommended that maintenance inspection is done once a year .
Maintenance inspection
• Check the hinges
necessary .
• Check the canvas
Disconnect mains before
removing the covers.
on the shower table and the safety rail
①
for damage and replace if necessary .
③
1
2
1
3
2
. Tighten if
②
en
en
13