Operation
62
Operating Instructions Cubis
Procedure
If the isoCAL function is set to automatic start and the isoCAL function is
t
triggered:
The [isoCAL] button flashes in the operating display.
y
Wait until the isoCAL function is executed.
t
In the operating display, a time display counts down from 15 seconds
y
to 0.
If no load change or no operation takes place on the device before
y
the expiration of the time display: The isoCAL function starts.
If the isoCAL function is set to manual start and the isoCAL function is
t
triggered:
The [isoCAL] button flashes in the operating display.
y
Press the [isoCAL] button.
t
The isoCAL function starts.
y
If the isoCAL function is complete: The device confirms the completion
y
of the calibration and adjustment process with an acoustic signal, and
the calibration report is displayed.
To output the calibration report via a connector: Press the [Print
t
Memory] button.
To close the calibration report and return to the previous display: Press
t
the [OK] button.
8.5.2
Internally Calibrating and Adjusting the Device
Requirement
The weighing pan is unloaded.
Procedure
Open the main menu.
t
Select the "Internal Adjustment" task.
t
The internal calibration and adjustment function is executed.
y
If automatic leveling is configured: The device levels itself automatically.
y
If the calibration and adjustment function is complete: The device
y
confirms the completion of the calibration and adjustment process with
an acoustic signal, and the calibration report is displayed.
To output the calibration report via a connector: Press the [Print
t
Memory] button.
To close the calibration report and return to the previous display: Press
t
the [OK] button.
| MCA Models
®