The Midi Pro kit comprises
(a) Midi Pro pipette controller
(b) Charger; either
220V-240V 50/60Hz 3-pin (UK),
100V-120V 50/60Hz 2-pin (USA),
220V-240V 50/60Hz 2-pin (Australia)
(c) Two spare hydrophobic filters, 1x 0.45µm and 1x 0.20µm
(d) Bench stand / wall bracket
(e) Instruction manual
Safety Instructions
The Midi Pro has been designed with safety in mind. They are lightweight and balance safely in the
hand. The triggers have been carefully designed to eliminate pressure points and control of intake and
delivery is therefore made easier.
It is recommended when using deleterious liquids that the user wears protective clothing including
goggles and gloves. If in doubt ask your safety officer. Please consult and establish appropriate health
and safety practices applicable to the liquid being dispensed within your working environment.
WARNING!
• Only use this instrument within the recommended limits of its chemical resistance.
• Never use with fuming liquids which are incompatible with ASA, ABS, silicone or polyproplyene.
• Never use with highly flammable liquids.
• Observe the reagent manufacturer's safety information.
• Never use the instrument in an atmosphere with danger of explosion.
• When not in use store on bench stand or wall bracket.
• Never use force when in use.
• Should the instrument fail to work to specification stop pipetting and check instructions
headed Trouble Shooting (page 7).
• Only recharge with manufacturer's original charger.
• All repairs must be carried out by an authorized dealer.
• Only use the manufacturer's accessories and spare parts.
• Charger must only be used indoors and protected against moisture.
4
Midi Pro
Operator Instructions
220V-240V 50/60Hz 2-pin (Europe),
100V-120V 50/60Hz 2-pin (Japan),
Recharging the Midi Pro
Before starting to operate, carefully check mains voltage and recharge batteries.
To charge the Midi Pro
1 Recharging temperature +10°C to +35°C.
2 Plug the charger jack (6) into the base of the Midi Pro.
3 Plug the charger (5) into a suitable wall socket. The red indicator on the charger will light up.
4 Place Midi Pro onto bench stand.
5 Allow unit to charge for 14 hours from empty to fully charged.
6 Once the Midi Pro is fully charged, the charger automatically switches to a trickle charge operation
and the red indicator will turn off. If the charge to the Midi Pro is interrupted and then re-started, the
charger starts its full 14 hour charging cycle again.
This will not damage the battery.
Only use the charger indoors and protect against moisture.
Fitting the Midi Pro wall bracket
The Midi Pro bench stand is supplied ready to use as a wall bracket. When used as a bench stand it is
recommended that the wall fitting is removed.