Fuse Replacement; Expected Service Life - Wellell Optima Turn Manual De Instrucciones

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 22
Optima Turn/English
7. Maintenance
7.1. General
(1).
Check main power cord and plug if there is abrasions or excessive wear.
(2).
Check mattress cover for signs of wear or damage. Ensure mattress cover and tubes are
stubbed together correctly.
(3).
Check airflow from the opening of the tubes in the quick connector. The airflow should
alternately blow every half of the cycle time between two tubes at the lower right corner
and one of the the other two tubes at the lower left corner will blow to give desired turning
angle when it is under turning mode.
(4).
Check the air hoses if there is kink or break. For replacement, please contact local
representative or place of purchase.

7.2. Fuse Replacement

(1).
Disconnect the plug from mains power when a
blown fuse is suspected.
(2).
Remove the cover of the fuse holder by means
of a flat screwdriver.
(3).
Insert a new fuse of the correct rating in, and
replace the cover of the fuse holder back. The
fuse should be rated as T1AL/250V type and
VDE approved.
7.3. Air Filter Replacement
(1).
Open the air filter cap located at the back of pump.
(2).
Replace with a clean filter and put the cover lid back.
The filter is reusable and can be washed gently with
a mild detergent and water. Dry the filter before use.
(3).
Suggested to check and replace air filter regularly
every six month.

8. Expected Service Life

The products are intended to offer safe and reliable operation when use or installed according
to the instructions provided by Apex Medical. Apex Medical recommends that the system be
inspected and serviced by authorized technicians if there are any signs of wear or concerns
with device function and indication on products. Otherwise, service and inspection of the
devices generally should not be required.