Inserting The Battery Into The Ultravision™ Generator; Operation Of The Ultravision™ Generator Using The Optional Ultravision™ Mains Converter; Inserting The Mains Converter Into The Ultravision™ Generator - alesi ULTRAVISION Manual De Usuario

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

Be aware
If the power supply units supplied with the recharging station does not have a suitable face plate that works with your local
power supply, please contact Alesi Surgical or your local distributor.
The recharging station includes temperature-compensated characteristics and has been designed for optimal charging and
life preservation of the batteries supplied with your Ultravision system. The use of other recharging stations even if mechan-
ically compatible is likely to dramatically shorten the operating life of your Ultravision system batteries. Do not use batteries
other than those supplied by Alesi Surgical with the recharging station.
Only use the power supply unit provided with the battery recharging station. The use of other power supply units may dam-
age batteries and recharging station.
Do not mechanically shock or drop a battery. This may damage or reduce the life of the battery.
Do not use the battery to power equipment other than the Ultravision generator. This may damage the battery and equip-
ment.
When the power indicator is red there is a fault with the recharger that may damage the batteries and the recharger should
not be used.
It is recommended that the battery is removed from the generator if the generator is not to be used for some time.
Advice
To maximise battery life, batteries should be returned to the recharging station at the end of use, such as at the end of each
day, and should be stored in as cool an environment as is possible. Leaving a battery uncharged over a weekend or holiday
break is not advised.
Do not discharge battery completely. This will greatly shorten battery life.
Inserting the battery into the Ultravision™ generator
To insert a battery into the generator:
1. Locate the battery slot on the right hand side of the Ultravision generator.
2. Ensure that the battery bay is empty by pulling the battery ejection trigger towards the front of the generator.
3. Ensure that the battery is oriented with the battery terminals closest to the generator, with the arrow facing up-
wards and pointing to the generator.
4. Insert the battery until it engages with the contact terminals in the generator.
5. Switch on power to the generator by pressing the on switch located on the front panel.
6. Switch off the device until set up is complete.
Caution
Ensure that battery is always inserted into the Ultravision generator and battery recharging station in the correct orientation.
Failure to do so may damage the battery and equipment and may cause a fire hazard.
Operation of the Ultravision™ Generator using the optional Ultravision™
Mains Converter
Inserting the Mains Converter into the Ultravision™ generator
To insert the mans converter into the generator:
1. Plug the power supply of the Mains Converter into the appropriate mains socket (this may require the use of one
of the regional adapters supplied with the Mains Converter). If the power supply units supplied with the Mains
Converter does not have a suitable face plate that works with your local power supply, please contact Alesi Surgi-
cal or your local distributor.
2. Insert the jack at the end of the power supply unit into the Mains Converter inlet socket.
3. Locate the battery slot on the right hand side of the Ultravision generator.
4. Ensure that the battery bay is empty by pulling the battery ejection trigger towards the front of the generator.
5. Ensure that the Mains Converter is oriented with the terminals closest to the generator, with the arrow facing
upwards and pointing to the generator.
6. Insert the Mains Converter until it engages with the contact terminals in the generator.
7. Turn on the mains power switch.
8. Turn the Ultravision generator fascia switch on when required and confirm that the LED on the generator fascia
16
Tabla de contenido
loading

Tabla de contenido